| Perma-Doc
Phantom - Catalog # 9404 |
| COMPATIBLE
WITH VariSource® / MicroSelectron® / GammaMed®
HDR Afterloading Systems. |
 |
| The
“PermaDoc Phantom” is shown
with the permanently embedded tungsten
wire centimeter scale and removable 3mm
diameter tandems. |
|
|
"
Your Daily Quality Assurance performed
with Ease and Confidence! "
|
 |
| Features: |
|
Verifies
Source Positioning and Stepping Accuracy |
|
Images Centimeter
Scale Directly on Film |
|
Requires
Single Exposure Only |
|
Set-up Accomplished
in Seconds |
|
Eliminates
All Guess-Work |
|
Precision
Built |
|
Compatible
with All HDR Units |
|
Fulfills
Documentary QA Requirements |
For compatibility to other HDR afterloading systems,
unit-specific tandems are utilized. |
 |
| Description: |
 |
| The
"PERMA-DOC PHANTOM" provides the necessary
PERMANENT documentation required by the NRC's
new and stringent quality assurance tests associated
with HDR. The PERMA-DOC PHANTOM, with its centimeter
scale automatically transferred to the film by
fluorescent radiation, readily identifies up to
20 source positions and their stepping accuracy.
Two matching 3/8" plexi plates allow for
insertion of a 10" x 12" ready-pack
V-film. The film is exposed by the HDR Ir-192
source with dwell times of I second over a maximum
of 20 positions spaced 1 cm apart. The resulting
film displays the source images superimposed on
the image of the centimeter scale. |
| Source
Images Superimposed on the Image of Centimeter
Scale!
|
 |
 |
| Shown
is part of the V-film exposed with an HDR Ir-192
source for dwell times of I second over a distance
of 20cm. Source position and stepping accuracy
are verified by scale lines that pass exactly
through the centers of the corresponding source
images. Most importantly, the first two distally
located markers are spaced such as to represent
the distance from tip of tandem to center of active
source. For example: Tip of tandem to center of
source for the GammaMed 12i sources is 3.75mm,
i.e., separation of the first two tungsten wires
is 3.75mm. With the incorporation of the distal
marker wire, the necessity to prick a hole into
the film is therefore eliminated. |